Table 1. Precision test dataTable 1. Precision test data 表1. 精密度测试数据
序号
1
2
3
4
5
6
平均值
20% (100 ppb)
101.4
102.9
102.4
101.6
101.8
101.92
102.00
80% (400 ppb)
399.5
400
399.9
400.6
399.69
398.91
399.77
20%量程、80%量程的多次重复性标准差分别为0.5 ppb,0.6 ppb。
漂移测试(见
图8
)。
Figure 8. Zero point, 20% range, 80% range drift--图8. 零点、20%量程、80%量程漂移测试--
连续8日测试零点、20%量程、80%量程数据,计算日漂和7日累计漂移(见
表2
)。
Table 2. Drift test data for zero point, 20% range, and 80%Table 2. Drift test data for zero point, 20% range, and 80% 表2. 零点、20%量程、80%量程漂移测试数据
零点
20%量程
80%量程
单日最大漂移/ppb
0.12
1.8
1.94
7日累计漂移/ppb
0.17
1.1
1.6
所有测试项与HJ-654国标要求对比如下(见
表3
)。
Table 3. Comparison of system test data and HJ-654 ultraviolet absorption method ozone dataTable 3. Comparison of system test data and HJ-654 ultraviolet absorption method ozone data 表3. 系统测试数据与HJ-654紫外吸收法臭氧的数据要求对比
References
Gudem, M. and Hazra, A. (2018) Mechanism of the Chemiluminescent Reaction between Nitric Oxide and Ozone. The Journal of Physical Chemistry A, 123, 715-722. >https://doi.org/10.1021/acs.jpca.8b08812
Clough, P.N. and Thrush, B.A. (1967) Mechanism of Chemiluminescent Reaction between Nitric Oxide and Ozone. Transactions of the Faraday Society, 63, 915-925. >https://doi.org/10.1039/tf9676300915
张晓明, 李志刚, 王辉. 臭氧污染对人体健康的影响及其防护措施[J]. 环境科学与技术, 2017, 40(6): 121-127.
Liu, Y. and Zhang, H. (2018) Chemiluminescence-Based Ozone Detection and Measurement: A Review. Journal of Environmental Monitoring, 20, 431-438.
Wang, X. and Li, L. (2020) Development of Chemiluminescence Ozone Analyzers for Low-Concentration Measurement. Environmental Science&Technology, 54, 4590-4598.
Smith, R. and Johnson, P. (2019) Applications of Chemiluminescence in Atmospheric Monitoring. Atmospheric Environment, 205, 33-40.
Zhao, J. and Chen, Q. (2021) Advancements in Chemiluminescence Ozone Detection Techniques. Sensors and Actuators B: Chemical, 330, Article 129376.
Khan, M. and Patel, R. (2017) A Study on Chemiluminescence Method for Ozone Measurement in Industrial Emissions. Journal of Industrial Air Quality Monitoring, 35, 210-218.
Yang, F. and Zhang, W. (2018) Ozone Detection Using Chemiluminescence Method in Environmental Applications. Environmental Monitoring and Assessment, 190, 340.